B&K Precision 4045B - Manual User Manual

Page 104

background image

104

Range

SIN: 16 to 1000; divisible by 4

SQU: 2 to 1000; divisible by 2

TRI:

16 to 1000; divisible by 4

NOIS: 16 to 1000,

ANO: 16 to 1000,

Rounding:

to integer value.

Scale

Type:

Numeric. MIN sets the scale to 1;

MAX sets the scale to 100

Range:

1 to 100 (See considerations)

Rounding;

to integer value.

Command Type: Setting only

Setting

Syntax:

:ARBitrary:PREDefined

, , ,

Examples:

:ARB:PRED SIN,1,1e3,100

Considerations:

1) The start address and the length must meet the

specification that. Start address + Length - 1 <= 1,000

2) The 'scale' refers to the scaling of the waveform as a

percentage of full scale. A scale of 100% will, under the

correct conditions, generate a waveform whose data

values range from -2047 to +2047. These 'correct

conditions' are set by the 'offset' value. This offset is the

value of the data at the start address, and determines

the maximum scale settable. The following table shows

the data values required in order to achieve maximum

scale.

SHAPE

DATA

SIN

0

SQU

0