85

FREQ 5E3

FREQ MAXIMUM

FREQ MIN

Query

Syntax: FREQuency?[MAXimum|MINimum]

Examples: FREQ?

FREQ? MAX

Response: NR3

Considerations:

1) The MIN and MAX arguments refer to currently

settable minimum or maximum.

2) FIXed is alias for CW.

Point Rate

RATE

This command is used to set the point rate. It is

coupled with the frequency of the waveform by the

relation:

Frequency = 1/(Point Rate * Wavelength)

Thus changing the point rate will result in a change

in frequency.

Arguments

Type: Numeric

Units: s, ms, us, ns

Range: 20ns to 50s