General technical specifications, Rf generator – Atec Rohde-Schwarz-CMW500 User Manual
Page 8
Version 11.00, May 2013
8
Rohde & Schwarz R&S
®
CMW500 Wideband Radio Communication Tester
General technical specifications
RF generator
Frequency range
70 MHz to 3300 MHz,
up to 6000 MHz with the
R&S
®
CMW-KB036 option
Frequency resolution
0.1
Hz
Frequency uncertainty
same as timebase + frequency resolution
Output level range
RF1 COM, RF2 COM
70 MHz to 100 MHz
continuous wave (CW)
–130 dBm to –15 dBm
peak envelope power (PEP)
up to –15 dBm
overranging (PEP)
up to –10 dBm
100 MHz to 3300 MHz
continuous wave (CW)
–130 dBm to –5 dBm
peak envelope power (PEP)
up to –5 dBm
overranging (PEP)
up to 0 dBm
3300 MHz to 6000 MHz
continuous wave (CW)
–120 dBm to –15 dBm
peak envelope power (PEP)
up to –15 dBm
overranging (PEP)
up to –10 dBm
maximum input DC level
0 V DC
RF1 OUT
70 MHz to 100 MHz
continuous wave (CW)
–120 dBm to –7 dBm
peak envelope power (PEP)
up to –7 dBm
overranging (PEP)
up to +3 dBm
100 MHz to 3300 MHz
continuous wave (CW)
–120 dBm to +3 dBm
peak envelope power (PEP)
up to +3 dBm
overranging (PEP)
up to +13 dBm
3300 MHz to 6000 MHz
continuous wave (CW)
–110 dBm to –7 dBm
peak envelope power (PEP)
up to –7 dBm
overranging (PEP)
up to +3 dBm
maximum input DC level
0 V DC
Output level uncertainty
in temperature range +20 °C to +35 °C,
no overranging
RF1 COM, RF2 COM
output level > –120 dBm
70 MHz to 100 MHz
< 1.2 dB
1
100 MHz to 3300 MHz
< 0.6 dB
1
3300 MHz to 6000 MHz
< 1.2 dB
1
RF1 OUT
output level > –112 dBm
70 MHz to 100 MHz
< 1.6 dB
1
100 MHz to 3300 MHz
< 0.8 dB
1
3300 MHz to 6000 MHz
< 1.6 dB
1
Output level uncertainty
in temperature range +5 °C to +45 °C,
no overranging
RF1 COM, RF2 COM
output level > –120 dBm
70 MHz to 100 MHz
< 2.0 dB
1
100 MHz to 3300 MHz
< 1.0 dB
1
3300 MHz to 6000 MHz
< 2.0 dB
1
RF1 OUT
output level > –112 dBm
70 MHz to 100 MHz
< 2.0 dB
1
100 MHz to 3300 MHz
< 1.0 dB
1
3300 MHz to 6000 MHz
< 2.0 dB
1
1
Valid for a 12-month calibration interval.