Spectrum measurements, Power meter – Atec Rohde-Schwarz-CMW500 User Manual
Page 25
Version 11.00, May 2013
Rohde & Schwarz R&S
®
CMW500 Wideband Radio Communication Tester 25
Analysis modes
with the R&S
®
CMW-KM400 option
QPSK, WCDMA
with the R&S
®
CMW-KM400 and
R&S
®
CMW-KM401 options
WCDMA + HSDPA, WCDMA + HSUPA,
WCDMA + HSPA
with the R&S
®
CMW-KM400 and
R&S
®
CMW-KM401 and
R&S
®
CMW-KM403 options
WCDMA + HSPA+
Measured parameters
numeric results and standard deviation
error vector magnitude (EVM),
magnitude error (ME),
phase error (PE),
frequency error,
I/Q origin offset,
I/Q imbalance,
UE power,
power steps,
phase discontinuity,
CDP,
CDE
graphical
EVM versus time, EVM versus chip,
ME versus time, ME versus chip,
PE versus time, PE versus chip,
FE versus time,
UE versus time,
PS versus slot,
PD versus slot,
CDP versus slot,
CDE versus slot,
CD monitor
Error vector magnitude (EVM)
Measurement range
up to 25 %, RMS
Inherent EVM
< 2.5 %, RMS
Measurement length
half-slot, 1 slot, multislot (1 to 120)
Frequency error
Measurement range
±3 kHz
Frequency measurement uncertainty
< 20 Hz + drift of timebase,
see general technical specifications
I/Q origin offset
Inherent I/Q offset
for average ≥ 10 measurements
< –55 dB
I/Q imbalance
Inherent I/Q imbalance
< –50 dB
Spectrum measurements
Adjacent channel leakage ratio
RMS detector
Filter
3.84 MHz, RRC, α = 0.22, WCDMA filter
Dynamic range
first adjacent channel at ±5 MHz
> 54 dB
second adjacent channel at ±10 MHz
> 57 dB
Expected nominal power setting for full
dynamic range
RF1 COM, RF2 COM
–4 dBm to +42 dBm
13
Uncertainty
for –33 dBc first adjacent channel level
< 0.5 dB
for –43 dBc second adjacent channel level < 0.5 dB
Measurement length
1 slot (2560 chip)
Power meter
UE power measurement
RMS detector
Filter
bandpass, 6.3 MHz, RRC, α = 0.22
Level range
–55 dBm to +42 dBm
13
Level uncertainty
see general technical specifications
Measurement length
half-slot, 1 slot
13