beautypg.com

Spectrum measurements, Power meter – Atec Rohde-Schwarz-CMW500 User Manual

Page 25

background image

Version 11.00, May 2013

Rohde & Schwarz R&S

®

CMW500 Wideband Radio Communication Tester 25

Analysis modes

with the R&S

®

CMW-KM400 option

QPSK, WCDMA

with the R&S

®

CMW-KM400 and

R&S

®

CMW-KM401 options

WCDMA + HSDPA, WCDMA + HSUPA,
WCDMA + HSPA

with the R&S

®

CMW-KM400 and

R&S

®

CMW-KM401 and

R&S

®

CMW-KM403 options

WCDMA + HSPA+

Measured parameters

numeric results and standard deviation

error vector magnitude (EVM),
magnitude error (ME),
phase error (PE),
frequency error,
I/Q origin offset,
I/Q imbalance,
UE power,
power steps,
phase discontinuity,
CDP,
CDE

graphical

EVM versus time, EVM versus chip,
ME versus time, ME versus chip,
PE versus time, PE versus chip,
FE versus time,
UE versus time,
PS versus slot,
PD versus slot,
CDP versus slot,
CDE versus slot,
CD monitor

Error vector magnitude (EVM)

Measurement range

up to 25 %, RMS

Inherent EVM

< 2.5 %, RMS

Measurement length

half-slot, 1 slot, multislot (1 to 120)

Frequency error

Measurement range

±3 kHz

Frequency measurement uncertainty

< 20 Hz + drift of timebase,
see general technical specifications

I/Q origin offset

Inherent I/Q offset

for average ≥ 10 measurements

< –55 dB

I/Q imbalance

Inherent I/Q imbalance

< –50 dB

Spectrum measurements

Adjacent channel leakage ratio

RMS detector

Filter

3.84 MHz, RRC, α = 0.22, WCDMA filter

Dynamic range

first adjacent channel at ±5 MHz

> 54 dB

second adjacent channel at ±10 MHz

> 57 dB

Expected nominal power setting for full
dynamic range

RF1 COM, RF2 COM

–4 dBm to +42 dBm

13

Uncertainty

for –33 dBc first adjacent channel level

< 0.5 dB

for –43 dBc second adjacent channel level < 0.5 dB

Measurement length

1 slot (2560 chip)

Power meter

UE power measurement

RMS detector

Filter

bandpass, 6.3 MHz, RRC, α = 0.22

Level range

–55 dBm to +42 dBm

13

Level uncertainty

see general technical specifications

Measurement length

half-slot, 1 slot



13

The maximum permissible continuous power is +34 dBm due to thermal limits.