beautypg.com

Audio specifications (prerequisite: r&s, Cmw-b400b), Audio analyzer – Atec Rohde-Schwarz-CMW500 User Manual

Page 18: Audio generator

background image

Version 11.00, May 2013

18

Rohde & Schwarz R&S

®

CMW500 Wideband Radio Communication Tester

Audio specifications (prerequisite: R&S

®

CMW-B400B)

Audio analyzer

Statistics

Statistical count

1 to 1000

Values

current, average, extremum, standard
deviation

Audio filter, weighting

Lowpass

off, 3 kHz, 4 kHz, 15 kHz

Highpass

off, 6 Hz, 50 Hz, 300 Hz

Weighting filter

off, A-weighting, CCITT, C-message

Connectors AF 1 IN / AF 2 IN

Input impedance

100 kΩ | 15 pF

AF voltmeter

Frequency range

50 Hz to 21 kHz

Level range

50 µV to 30 V

Level resolution

at level < 1 mV

1 µV

at level ≥ 1 mV

0.1 %

Level uncertainty

at 1mV ≤ level < 20 mV

< 1.3% + resolution

at 20 mV ≤ level ≤ 20 V

< 1.0 % + resolution

THD+N and THD meter

Measurement bandwidth

21

kHz

Frequency range

100 Hz to 10.5 kHz

Level range

10 mV to 30 V

Resolution

0.01

%

Inherent distortion THD+N

at 100 mV ≤ level ≤ 20 V

< 0.05 %

Inherent distortion THD

at 100 mV ≤ level ≤ 20 V

< 0.025 %

Uncertainty

at 100 mV ≤ level ≤ 2 V

< 1 % + inherent distortion

at 2 V ≤ level ≤ 20 V

< 2 % + inherent distortion

Audio generator

Connectors AF 1 OUT / AF 2 OUT

Output impedance

< 4 Ω

Maximum output current

20 mA

AF sine generator

Frequency range

20 Hz to 21 kHz

Frequency uncertainty

same as time base ,
see base unit specifications

Frequency resolution

1 Hz

Level range

10 µV to 5 V

Level resolution

at level < 10 mV

10 µV

at level ≥ 10 mV

0.1 %

Level uncertainty

at level ≥ 1 mV and
frequency ≤ 10 kHz

≤ 1.5 % + resolution

THD+N

at level ≥ 100 mV,
into load ≥ 600 Ω,
measurement bandwidth: 21.9 kHz

≤ 0.05 %

THD

at level ≥ 100 mV,
into load ≥ 600 Ω,
measurement bandwidth: 21.9 kHz