Power meter – Atec Rohde-Schwarz-CMW500 User Manual
Page 11
Version 11.00, May 2013
Rohde & Schwarz R&S
®
CMW500 Wideband Radio Communication Tester 11
Trigger
Trigger sources
BASE: external TRIG A,
BASE: external TRIG B,
GPRF: free run,
GPRF: IF power,
BB generators,
BB signaling
Power meter
Frequency range
70 MHz to 3300 MHz,
up to 6000 MHz with the
R&S
®
CMW-KB036 option
Frequency resolution
0.1
Hz
Resolution bandwidths
Gaussian, 1 kHz to 10 MHz, in 1/3/5 steps,
bandpass, 1 kHz to 30 MHz, in 1/3/5
steps, RRC, α = 0.1,
3.84 MHz, RRC, α = 0.22, WCDMA filter,
1.2288 MHz, CDMA filter
Expected nominal power setting range
for ADC full scale
RF1 COM, RF2 COM
70 MHz to 100 MHz
–37 dBm to +42 dBm
2
100 MHz to 3300 MHz
–47 dBm to +42 dBm
2
3300 MHz to 6000 MHz
–37 dBm to +42 dBm
2
Level range
RF1 COM, RF2 COM
70 MHz to 100 MHz
continuous power (CW)
–74 dBm
3
to +34 dBm
peak envelope power (PEP)
up to +42 dBm
2
100 MHz to 3300 MHz
continuous power (CW)
–84 dBm
3
to +34 dBm
peak envelope power (PEP)
up to +42 dBm
2
3300 MHz to 6000 MHz
continuous power (CW)
–74 dBm
3
to +34 dBm
peak envelope power (PEP)
up to +42 dBm
2
maximum input DC level
0 V DC
Level uncertainty
in temperature range +20 °C to +35 °C
RF1 COM, RF2 COM
70 MHz to 100 MHz
< 1.0 dB
4
100 MHz to 3300 MHz
< 0.5 dB
4
3300 MHz to 6000 MHz
< 1.0 dB
4
Level uncertainty
in temperature range +5 °C to +45 °C
RF1 COM, RF2 COM
70 MHz to 100 MHz
< 1.2 dB
4
100 MHz to 3300 MHz
< 0.7 dB
4
3300 MHz to 6000 MHz
< 1.2 dB
4
Level linearity with fixed expected
nominal power setting
in temperature range +20 °C to +35 °C
RF1 COM, RF2 COM
level range 0 dB to –40 dB
< 0.15 dB, typ. < 0.1 dB
Level resolution
0.01
dB
Level repeatability
typical values after 1 h warm-up time,
always returning to same level and
frequency, no temperature change,
insignificant time change
input level ≥ –40 dBm
< 0.01 dB
input level < –40 dBm
< 0.03 dB
2
The maximum permissible continuous power is +34 dBm due to thermal limits.
3
RBW → 1 kHz.
4
Valid for a 12-month calibration interval.