beautypg.com

Spectrum measurements – Atec Rohde-Schwarz-CMW500 User Manual

Page 36

background image

Version 11.00, May 2013

36

Rohde & Schwarz R&S

®

CMW500 Wideband Radio Communication Tester

Frequency error

Measurement range

±80 kHz

Frequency measurement uncertainty

< 20 Hz

19

+ drift of timebase,

see general technical specifications

Spectrum measurements

Adjacent channel leakage ratio

Filter

E-UTRA

rectangle 1.08 MHz, 2.7 MHz, 4.5 MHz,
9 MHz, 13.5 MHz, 18 MHz

UTRA

1.28 MHz, RRC, α = 0.22, WCDMA filter
3.84 MHz, RRC, α = 0.22, WCDMA filter
7.68 MHz, RRC, α = 0.22, WCDMA filter

Dynamic range

E-UTRA, f

RF

< 3300 MHz

> 50 dB

E-UTRA, f

RF

> 3300 MHz

> 45 dB

UTRA128, f

RF

< 3300 MHz

> 52 dB

UTRA384, f

RF

< 3300 MHz

> 52 dB

UTRA768, f

RF

< 3300 MHz

> 49 dB

UTRA128, f

RF

> 3300 MHz

> 49 dB

UTRA384, f

RF

> 3300 MHz

> 49 dB

UTRA768, f

RF

> 3300 MHz

> 46 dB

Expected nominal power setting for full
dynamic range

RF1 COM, RF2 COM

see general technical specifications

Spectrum emission mask (SEM)

Frequency span

70 MHz

Noise floor

RBW → ≤ 1 MHz, expected nominal
power < 12dBm, f

RF

< 3300 MHz

< –58 dBm

RBW → ≤ 1 MHz, expected nominal
power < 12dBm, f

RF

> 3300 MHz

< –55 dBm

19

For frequency error measurements relative to the downlink signal of the same R&S

®

CMW500, the uncertainty is in line with 3GPP TS 36.521-1 V9.3.0.