Modulation analysis, Code domain, Spectrum measurements – Atec Rohde-Schwarz-CMW500 User Manual
Page 57
Version 11.00, May 2013
Rohde & Schwarz R&S
®
CMW500 Wideband Radio Communication Tester 57
Modulation analysis
Filter
1.28 MHz, RRC, α = 0.22, TD-SCDMA
filter
Level range
–28 dBm to +42 dBm
43
Analysis modes
TD-SCDMA uplink
QPSK, DPCH,
QPSK, DPCH + HSDPA,
16QAM, DPCH + HSUPA
Measured parameters
numeric results and standard deviation
error vector magnitude (EVM),
magnitude error (ME),
phase error (PE),
frequency error,
I/Q origin offset,
I/Q imbalance,
UE power
graphical
EVM versus time,
ME versus time,
PE versus time
Error vector magnitude (EVM)
Measurement range
up to 25 %, RMS
Inherent EVM
< 2.5 %, RMS
Measurement length
1 slot, multislot (1 to 112)
Frequency error
Measurement range
±3 kHz
Frequency measurement uncertainty
< 20 Hz + drift of timebase,
see general technical specifications
I/Q origin offset
Inherent I/Q offset
for average ≥ 10 measurements
< –55 dB
I/Q imbalance
Inherent I/Q imbalance
< –50 dB
Code domain
Filter
1.28 MHz, RRC, α = 0.22, TD-SCDMA
filter
Level range
–28 dBm to +42 dBm
43
Analysis modes
TD-SCDMA uplink
QPSK, DPCH,
QPSK, DPCH + HSDPA,
16QAM, DPCH + HSUPA
Measured parameters
numeric results and standard deviation
code domain error (CDE)
graphical
code domain power versus code
Code domain power versus code
Uncertainty
< 0.4 dB
Measurement length
1 slot
Code domain error (CDE)
Uncertainty
< 0.4 dB
Measurement length
1 slot
Spectrum measurements
Adjacent channel leakage ratio
RMS detector
Filter
1.28 MHz, RRC, α = 0.22, TD-SCDMA
filter
Dynamic range
first adjacent channel at ±1.6 MHz
> 53 dB
second adjacent channel at ±3.2 MHz
> 61 dB
Expected nominal power setting for full
dynamic range
RF1 COM, RF2 COM
–3 dBm to +42 dBm
44
43
The maximum permissible continuous power is +34 dBm due to thermal limits.