beautypg.com

Atec Rohde-Schwarz-CMW500 User Manual

Page 69

background image

Version 11.00, May 2013

Rohde & Schwarz R&S

®

CMW500 Wideband Radio Communication Tester 69

Mechanical resistance

non-operating mode

Vibration

sinusoidal

in line with EN 60068-2-6,
5 Hz to 150 Hz, 0.15 mm amplitude const.,
55 Hz to 150 Hz, 0.5 g const.

Vibration

random

in line with EN 60068-2-64,
10 Hz to 300 Hz, acceleration 1.2 g RMS

Shock

in line with MIL-STD-810E, method 516.4,
procedure I, 40 g shock spectrum

Power rating

Rated voltage

100 V to 240 V AC (± 10 %)

Rated frequency

50 Hz to 60 Hz (± 5 %)

Rated power

max. 850 VA

standby

max. 13 W

Power consumption

R&S

®

CMW500 and R&S

®

CMW270

single tester, non-signaling

approx. 200 W

R&S

®

CMW280 single tester, non-signaling approx. 180 W

Display

selected with R&S

®

CMW-S600B or

R&S

®

CMW-S600D or R&S

®

CMW-S600F

configuration
(front panel with display and keypad)

Size

21 cm TFT color display (8.4")

Resolution

800 × 600 pixel (SVGA resolution)

Pixel failure rate

< 1.1 × 10

–5

Dimensions of the R&S

®

CMW500 or

R&S

®

CMW270

W × H × D, overall

465.1 mm × 197.3 mm × 517.0 mm
(18.31 in × 7.77 in × 20.35 in)

for rackmounting

19" 1/1, 4 HU, 450

Dimensions of the R&S

®

CMW280

W × H × D, overall

465.1 mm × 197.3 mm × 417.0 mm
(18.31 in × 7.77 in × 16.41 in)

for rackmounting

19" 1/1, 4 HU, 350

Weight of the R&S

®

CMW500 or

R&S

®

CMW270

single tester, non-signaling

approx. 14 kg
(approx. 31 lb)

with typical options

approx. 18 kg
(approx. 40 lb)

Weight of the R&S

®

CMW280

single tester, non-signaling

approx. 12 kg
(approx. 27 lb)

with typical options

approx. 14 kg
(approx. 31 lb)

Calibration interval

12 months

recommended for highest accuracy,
see specified RF generator and
RF analyzer level uncertainty

24 months

add 0.2 dB to specified RF generator and
RF analyzer level uncertainty