beautypg.com

Lte fdd enodeb rf analyzer (r&s, Cmw-kn500 option), Spectrum measurements – Atec Rohde-Schwarz-CMW500 User Manual

Page 30

background image

Version 11.00, May 2013

30

Rohde & Schwarz R&S

®

CMW500 Wideband Radio Communication Tester

Spectrum measurements

Adjacent channel leakage ratio

Filter

E-UTRA

rectangle 1.08 MHz, 2.7 MHz, 4.5 MHz,
9 MHz, 13.5 MHz, 18 MHz

UTRA

3.84 MHz, RRC, α = 0.22, WCDMA filter

Dynamic range

E-UTRA, f

RF

< 3300 MHz

> 45 dB

E-UTRA, f

RF

> 3300 MHz

> 42 dB

UTRA, f

RF

< 3300 MHz

> 52 dB

UTRA, f

RF

> 3300 MHz

> 49 dB

Expected nominal power setting for full
dynamic range

RF1 COM, RF2 COM

see general technical specifications

Spectrum emission mask (SEM)

Frequency span

70 MHz

Noise floor

RBW → 1 MHz, f

RF

< 3300 MHz

< –35 dBm

RBW → 1 MHz, f

RF

> 3300 MHz

< –32 dBm

RBW → 100 kHz, f

RF

< 3300 MHz

< –40 dBm

RBW → 100 kHz, f

RF

> 3300 MHz

< –37 dBm

RBW → 30 kHz, f

RF

< 3300 MHz

< –45 dBm

RBW → 30 kHz, f

RF

> 3300 MHz

< –42 dBm

LTE FDD eNodeB RF analyzer (R&S

®

CMW-KN500 option)

Bandwidth

1.4 MHz, 3 MHz, 5 MHz, 10 MHz, 15 MHz,
20 MHz

Frequency range

E-UTRA band 1, FDD

2110 MHz to 2170 MHz

E-UTRA band 2, FDD

1930 MHz to 1990 MHz

E-UTRA band 3, FDD

1805 MHz to 1880 MHz

E-UTRA band 4, FDD

2110 MHz to 2155 MHz

E-UTRA band 5, FDD

869 MHz to 894 MHz

E-UTRA band 6, FDD

875 MHz to 885 MHz

E-UTRA band 7, FDD

2620 MHz to 2690 MHz

E-UTRA band 8, FDD

925 MHz to 960 MHz

E-UTRA band 9, FDD

1844.9 MHz 1879.9 MHz

E-UTRA band 10, FDD

2110 MHz to 2170 MHz

E-UTRA band 11, FDD

1475.9 MHz to 1495.9 MHz

E-UTRA band 12, FDD

729 MHz to 746 MHz

E-UTRA band 13, FDD

746 MHz to 756 MHz

E-UTRA band 14, FDD

758 MHz to 768 MHz

E-UTRA band 15, FDD

2600 MHz to 2620 MHz

E-UTRA band 16, FDD

2585 MHz to 2600 MHz

E-UTRA band 17, FDD

734 MHz to 746 MHz

E-UTRA band 18, FDD

860 MHz to 875 MHz

E-UTRA band 19, FDD

875 MHz to 890 MHz

E-UTRA band 20, FDD

791 MHz to 821 MHz

E-UTRA band 21, FDD

1495.9 MHz to 1510.9 MHz

E-UTRA band 22, FDD,
prerequisite: R&S

®

CMW-KB036 option

3510 MHz to 3590 MHz

E-UTRA band 23, FDD

2180 MHz to 2200 MHz

E-UTRA band 24, FDD

1525 MHz to 1559 MHz

E-UTRA band 25, FDD

1930 MHz to 1995 MHz

Level setting

manual

mode

Statistics

Statistical count

1 to 1000

Values

current, average, minimum/maximum,
standard deviation

Trigger

Trigger source

BASE: external TRIG A,
BASE: external TRIG B,
GPRF: BB generator,
LTE: free run (fast sync),
LTE: IF power