Power measurement

Slot power

RMS detector

Level range

–50 dBm to +30 dBm, RMS

Level uncertainty

see general technical specifications

Modulation analysis

Measured parameters

numeric results

error vector magnitude (EVM),
magnitude error (ME),
phase error (PE),
frequency error,
I/Q origin offset,
TX power,
peak power,
reference symbol power,
OFDM symbol power

graphical

EVM versus OFDMA symbol,
ME versus OFDMA symbol,
PE versus OFDMA symbol,
OFDM symbol power
spectrum flatness,
spectrum ACLR
I/Q constellation

Error vector magnitude (EVM)

Inherent EVM

for average ≥ 10 subframes

–20 dBm ≤ input level < +30 dBm,
RMS

< 1.5 %, RMS

Frequency error

Measurement range

±80 kHz

Frequency measurement uncertainty

< 20 Hz

16

+ drift of timebase,

see general technical specifications

Spectrum measurements

Adjacent channel leakage ratio

Filter

E-UTRA

rectangle 1.08 MHz, 2.7 MHz, 4.5 MHz,
9 MHz, 13.5 MHz, 18 MHz

UTRA

1.28 MHz, RRC, α = 0.22, WCDMA filter
3.84 MHz, RRC, α = 0.22, WCDMA filter
7.68 MHz, RRC, α = 0.22, WCDMA filter

Dynamic range

E-UTRA, f

RF

< 3300 MHz

> 50 dB

E-UTRA, f

RF

> 3300 MHz

> 45 dB

UTRA, f

RF

< 3300 MHz

> 52 dB

UTRA, f

RF

> 3300 MHz

> 49 dB

Expected nominal power setting for full
dynamic range

RF1 COM, RF2 COM

see general technical specifications

Spectrum emission mask (SEM)

Frequency span

70 MHz

Noise floor

RBW → ≤ 1 MHz, expected nominal
power < 12dBm, f

RF

< 3300 MHz

< –58 dBm

RBW → ≤ 1 MHz, expected nominal
power < 12dBm, f

RF

> 3300 MHz

< –55 dBm

16

For frequency error measurements relative to the downlink signal of the same R&S

®

CMW500, the uncertainty is in line with 3GPP TS 36.521-1 V9.3.0.