Code domain, Spectrum measurements, Power meter – Atec Rohde-Schwarz-CMW500 User Manual
Page 61
Version 11.00, May 2013
Rohde & Schwarz R&S
®
CMW500 Wideband Radio Communication Tester 61
Frequency error
Measurement range
±3 kHz
Frequency measurement uncertainty
< 20 Hz + drift of timebase,
see general technical specifications
Carrier feedthrough
Inherent carrier feedthrough
for average ≥ 10 measurements
< –55 dB
I/Q imbalance
Inherent I/Q imbalance
< –50 dB
Code domain
Filter
1.23 MHz, RRC, α = 0.22, CDMA filter
Level range
–28 dBm to +42 dBm
47
Measured parameters
numeric values of current, average,
max. and min. values
code domain power (CDP),
code domain error (CDE)
graphical
code domain power versus code,
code domain error versus code
Code domain power versus code
Uncertainty
< 0.4 dB
Measurement length
616 chip (0.5 ms)
Code domain error versus code
Measurement uncertainty
< 0.4 dB
Measurement length
616 chip (0.5 ms)
Spectrum measurements
Adjacent channel power
RMS detector, at the selected frequency
offsets
Filter
30 kHz Gaussian
Frequency offset interval
up to 10 adjacent channels on each side
–4 MHz to +4 MHz
Dynamic range
> 70 dB
Expected nominal power setting for full
dynamic range
RF1 COM, RF2 COM
+1 dBm to +42 dBm
47
Uncertainty
for –43 dBc adjacent channel level
< 0.5 dB
Measurement length
one power control group
1536 chip (1.25 ms)
Power meter
MS power
Filter narrowband
bandpass,
1.25
MHz
Filter
wideband
bandpass, 8 MHz
Level range
–55 dBm to +42 dBm
47
Level uncertainty
see general technical specifications
Measurement length
616 chip (0.5 ms)
47
The maximum permissible continuous power is +34 dBm due to thermal limits.