YSI Vertical Profiler Systems User Manual
Page 33

YSI Profiler System USER Manual
-----------------Main------------------ 
1-Run 5-System 
2-Calibrate 6-Report 
3-File 7-Sensor 
4-Status 8-Advanced 
 
Next, select 4-Status to make certain that the sonde is not set for logging. Item number seven 
should read Logging: Inactive. Check your sonde manual for details on turning off the logging 
activity, or see below during the final sonde menu checks. Normally the active logging 
parameter would not be set. 
 ----------------Status----------------- 
1-Version:2.25 
2-Date=11/17/05 
3-Time=16:24:36 
4-Bat volts:12.3 
5-Bat life 20.0 days 
6-Free bytes:295424 
7-Logging:Inactive 
 
Note that battery voltage for the sonde is 12.3 volts and has an expected battery life of more than 20 days 
based on parameters that are set. Although you see Date and Time in this screen, this does not necessarily 
mean that Report setup has date and time enabled. You confirmed that these parameters were turned off 
when you displayed the Report setup menu above. 
 
Next, type 0 to return to Main menu. 
 
-----------------Main------------------ 
1-Run 5-System 
2-Calibrate 6-Report 
3-File 7-Sensor 
4-Status 8-Advanced 
 
Type 8-Advanced, to view a submenu that takes you into a variety of settings from calibration 
constants to timing of activities. In addition to checking some of the timing settings, it is 
important to verify that the Auto sleep function is enabled. This is one of the requirements of 
monitoring in the Profiler routine. 
 
---------------Advanced---------------- 
1-Cal constants 
2-Setup 
3-Sensor 
4-Data filter 
 
 
 
YSI Environmental
669523 Rev B
Page 3-4