9: Nonlinearity Correction, Stray Light Correction and Calibration

RAYSPHERE-1700-02-201304

53

Wavelength

range

400

nm –

500

nm

500

nm –

600

nm

600 nm –

700 nm

700 nm – 800

nm

800 nm – 900 nm

900 nm – 1100

nm

1.

Spec_Distr [%]

1.70

1.29

1.04

0.91

0.78

0.71

2.,1 Dark [%]

0.20 0.05

0.06

3.,1 T_TEC [%]

0.06

3.,2 StrayL [%]

ntia

Standard
Uncertainty [%]

1.71

1.29

1.04

0.91

0.78

0.71

Expanded
Uncertainty [%]
for k = 2

3.42

2.58

2.08

1.82

1.56

1.42

A missing value indicates that the value is less than 0.1 % or not known.

How to Estimate the Uncertainties for the Spectral Match

Test

An example illustrates the procedure for a measurement with a solar simulator. The following spectrum is
recorded.

RaySphere calculated the percentage of total irradiance in the wavelength range of 400 nm to1100 nm.