Chapter 7, Overview, Classifying the solar simulator – Ocean Optics RaySphere 1700 Install User Manual
Page 49: And displaying the results
RAYSPHERE-1700-02-201304
39
Chapter 7
Classifying the Solar Simulator
and Displaying the Results
Overview
RaySphere is designed for the spectral match test as described in detail in the international standard
IEC60904-9. For six wavelength intervals of interest, the percentage of total irradiance is specified in the
table shown below:
Interval Wavelength Range in nm Percentage of Total Irradiance in Wavelength Range 400 nm – 1100 nm
1
400 – 500
18.4%
2
500 – 600
19.9%
3
600 – 700
18.4%
4
700 – 800
14.9%
5
800 – 900
12.5%
6
900 – 1100
15.9%
The solar simulator requirements for classification for the spectral match test are as follows:
Classification
Spectral match to all intervals specified in the table above
A
0.75 – 1.25
B
0.6 – 1.4
C
0.4 – 2.0
The RaySphere software displays the results from spectral match test, the spectrum in absolute irradiance
mode, and the timing diagram with photocurrent in the software section Classification tab as shown in
the example below. See