beautypg.com

Chapter 7, Overview, Classifying the solar simulator – Ocean Optics RaySphere 1700 Install User Manual

Page 49: And displaying the results

background image

RAYSPHERE-1700-02-201304

39

Chapter 7

Classifying the Solar Simulator

and Displaying the Results

Overview

RaySphere is designed for the spectral match test as described in detail in the international standard
IEC60904-9. For six wavelength intervals of interest, the percentage of total irradiance is specified in the
table shown below:

Interval Wavelength Range in nm Percentage of Total Irradiance in Wavelength Range 400 nm – 1100 nm

1

400 – 500

18.4%

2

500 – 600

19.9%

3

600 – 700

18.4%

4

700 – 800

14.9%

5

800 – 900

12.5%

6

900 – 1100

15.9%

The solar simulator requirements for classification for the spectral match test are as follows:

Classification

Spectral match to all intervals specified in the table above

A

0.75 – 1.25

B

0.6 – 1.4

C

0.4 – 2.0

The RaySphere software displays the results from spectral match test, the spectrum in absolute irradiance
mode, and the timing diagram with photocurrent in the software section Classification tab as shown in
the example below. See

Classification Tab

for more information.