How to create a configuration file – Ocean Optics RaySphere 1700 Install User Manual
Page 47
6: Using Preconfigured Configuration Files
RAYSPHERE-1700-02-201304
37
Field
Meaning
Flash/continuous mode =>
Flash/Autotest =>
Internal/Manual
Internal/Auto
Operation mode: Auto
Choose Continuous/Valuefind or Autotest/Singlespec
Infos of the QE:
SN: QEB1351
Cal.file:E:QEB1351_20111213_OOIIrrad.cal
Delay[ms]: 19.99
It: 10
TEC[°C]: -12
Scan2Avr: 1
Boxcar: 5
Serial Number of the spectrometer. This number must be
correct.
Name of the calibration file and its location (information must
be correct).
Delay of the QE65000 spectrometer
Integration time of the QE65000 spectrometer in ms
(delimiter is “.”)
Detector temperature in °C
Scans to average to get one spectrum
Boxcar setting
NOTE:
Similar information is also displayed for the NIRQuest512 spectrometer for RaySphere Model 1700.
Straylight correction File:
E:QEB1351_EasyCorrM_.txt
The name of the stray light correction file and where it is
stored (both must be correct).
How to Create a Configuration File
► Procedure
1. Create a configuration file by pressing
or
in the System tab of the RaySphere software.
2. Follow the wizard to save data. All settings of the actual status are taken over into the
configuration file. A configuration file is in ASCII and can be edited manually.