Printronix Online Data Validator User Manual
Page 43

Scanning Beam Skew Adjustment
43
4.
Make sure the bar code is under the scanning beam.
5.
Press and hold the
RESET
button on the validator for 4 to 6
seconds until the READ light illuminates and the
CALIBRATION light begins flashing.
Figure 23. Calibrating the Validator
6.
Within a few seconds the CALIBRATION and READ light will
turn off. This indicates a successful calibration.
NOTE:
If the calibration light is still on or flashing, the calibration
has failed. In this case, refer to Table 2 on page 42 to
repeat the calibration procedure or to reset the validator
settings back to the last good calibration settings.
7.
Once you have a successful calibration, remove the calibration
bar code card.