Scanning Beam Skew Adjustment
43
4.
Make sure the bar code is under the scanning beam.
5.
Press and hold the
RESET
button on the validator for 4 to 6
seconds until the READ light illuminates and the
CALIBRATION light begins flashing.
Figure 23. Calibrating the Validator
6.
Within a few seconds the CALIBRATION and READ light will
turn off. This indicates a successful calibration.
NOTE:
If the calibration light is still on or flashing, the calibration
has failed. In this case, refer to Table 2 on page 42 to
repeat the calibration procedure or to reset the validator
settings back to the last good calibration settings.
7.
Once you have a successful calibration, remove the calibration
bar code card.
183467a
CALIBRATION
Light
RESET Button
Deck Lock
Lever
Calibration Bar
Code Card