2398 invalid dss type, 2399 invalid sync type – Quantum Data 801GC User Manual
Page 322

8-14
Chapter 8: Error Messages
Model 801GC, 801GF & 801GX¥Rev. A
2398
Invalid DSS type
An invalid digital separate sync type selection exists in the
file.
■
Reedit the digital separate sync type setting.
2399
Invalid sync type
An invalid sync select exists in the file.
■
Reedit the sync select setting.
2400
Digital separate sync type selection not
compatible with analog and/or digital sync
type
The generator does not support the given combination of
digital separate sync (H&V) selection with one or both of
the analog or digital composite sync type selections.
■
Select only one sync type.
2430
Vertical pulse delay extends sync beyond
blanking
The generator does not support vertical sync pulses with
any portion of the pulse occurring outside of the vertical
blanking period.
■
Decrease the vertical sync pulse width.
■
Decrease the vertical sync pulse delay.
■
Increase the vertical period (in lines).
■
Decrease the vertical active (in lines).
2450
Vertical pulse delay extends sync beyond
blanking
The generator does not support vertical sync pulses with
any portion of the pulse occurring outside of either vertical
blanking period with interlaced operation.
■
Decrease the vertical sync pulse width.
■
Decrease the vertical sync pulse delay.
■
Increase the vertical period (in lines).
■
Decrease the vertical active (in lines).