beautypg.com

Quantum Data 801GC User Manual

Page 273

background image

6-147

KNOB

rotate KNOB

Classification:

System parameter setting

Command Syntax:

KNOB

Limits:

1 = upper Format knob
2 = lower Image knob

positive integer = knob clicks in a clockwise direction
negative integer = knob clicks in a counterclockwise
direction

Query Syntax:

None

Description:

The KNOB command has the same affect on generator operation

as if an operator were manually turning either the “Format” or
“Image” knobs on the generator.

When a sequence is running, the “Format” knob loads sequences
from the directory selected by the SEQP command. Clockwise
rotation increases the index pointer for the directory. The “Image”
knob is used to go back and forth through the sequence steps.
Clockwise rotation selects higher step numbers.

Other Required Cmds:

None

Example:

K N O B 2 1 / / M o v e I m a g e k n o b 1 c l i c k
/ / C W

LCDS

LCD Status

Classification:

System parameter setting

Command Syntax:

None

Query Syntax:

LCDS?

Query Response:

Response in normal format/image selection mode…

H

=

V

=

Description:

The LCDS? query returns text string data that matches what is
shown in the LCD window of the generator.

Other Required Cmds:

None

Example:

L C D S ? / / R e t u r n t e x t s i m i l a r t o t e x t
/ / b e l o w

/ / H 3 2 1 5 = V G A _ m 3 < c r > < l f >

/ / V 6 0 5 1 = S M P T E 1 3 3