2330 vertical blanking too small, 2350 vertical pulse too small, 2370 vertical pulse too large – Quantum Data 801GC User Manual
Page 320: 2390 vertical pulse too large

8-12
Chapter 8: Error Messages
Model 801GC, 801GF & 801GX¥Rev. A
2321
Vertical active must be even when in inter-
laced scan mode
The generator does not support vertical active periods (in
lines) that cannot be evenly divided by two (2) when
interlaced operation is selected.
■
Make the vertical active period (in lines) evenly divisible
by 2.
■
Change to progressive (non-interlaced) scan mode.
2330
Vertical blanking too small
There must be at least 3 lines of vertical blanking with
interlaced operation.
■
Increase the vertical period.
■
Decrease vertical active.
■
Switch to non-interlaced operation.
2350
Vertical pulse too small
The generator does not support a vertical sync pulse width
value of less than one scan line in duration. The firmware
makes the pulse a half-line shorter than the entered value
when European type sync is selected.
■
Increase the vertical sync pulse width.
2370
Vertical pulse too large
The generator does not support a vertical sync pulse width
greater than the number of blanked scan lines per frame
with non-interlaced operation.
■
Decrease the value of vertical sync pulse width.
■
Increase the value of vertical period (in lines).
■
Decrease vertical active.
2390
Vertical pulse too large
The generator does not support a vertical sync pulse width
greater than the least number of blanked scan lines ((Vper
- Vact - 1)
÷
2 ) between fields with interlaced operation.
■
Decrease the value of vertical sync pulse width.
■
Increase the value of vertical period (in lines).
■
Decrease vertical active.