6-159
SCAN
SCAN fields per frame
Classification:
Format parameter setting
Command Syntax:
SCAN
Limits:
1 = progressive (non-interlaced)
2 = interlaced
Query Syntax:
SCAN?
Query Response:
Description:
The SCAN command establishes the number of fields scanned
per frame. Set to one (1) for progressive (non-interlaced) scan
and two (2) for interlaced scan. The SCAN? query returns the
current setting of SCAN.
Other Required Cmds:
The FMTU command instructs the generator to use the new
setting. The ALLU command updates hardware to the new setting
and redraws the test image.
Example:
S C A N / / S e l e c t i n t e r l a c e i n b u f f e r
F M T U / / U p d a t e h a r d w a r e t o c u r r e n t
/ / b u f f e r c o n t e n t s
SDLY
Sequence step DeLaY
Classification:
Sequence parameter setting
Command Syntax:
SDLY
Limits:
min = 0.0 seconds
max = 1.70E38 seconds (infinity for all practical purposes)
Query Syntax:
SDLY?
Query Response:
Description:
The SDLY command sets how long a sequence step will pause
before advancing to the next step in the Auto run mode. A
sequence step will use the last value set by the SDLY command.
The SDLY? query returns the current setting of SDLY.
Other Required Cmds:
SMOD setting must be equal to three (3) in order for the SDLY
setting to have any affect on sequence operation.
Example:
S D L Y 5 . 0 / / S e t d e l a y t o f i v e s e c o n d s
/ / p e r s t e p