Quantum Data 801GC User Manual
Page 277

6-151
MSIZ
light Meter SIZe
Classification:
System parameter setting
Command Syntax:
MSIZ
or
MSIZ
Limits:
Query Syntax:
MSIZ?
Query Response:
Description:
The MSIZ command establishes the physical size of the lightmeter
box(es) displayed in the BriteBox test image. The unit of measure
used is based on the current setting of the system level USIZ
parameter. It also affects the size of the cursor boxes in the
Persist image. Changing the size will not change the currently
displayed image. The MSIZ? query returns the current settings
of MSIZ based on the current setting of the system level USIZ
parameter.
Other Required Cmds:
The correct unit of measure should be selected with the USIZ
command prior to setting the size.
Example:
U S I Z 1 / / S e l e c t i n c h e s f o r u n i t s
M S I Z 2 . 0 / / S e t s i z e t o 2 . 0 i n c h e s I M G L
/ / B r i t e B o x S e l e c t B r i t e B o x t e s t
/ / i m a g e I M G U D r a w s e l e c t e d i m a g e
/ / u s i n g n e w s i z e
NAMF
NAMe Find
Classification:
Directory editor control
Command Syntax:
None
Query Syntax:
NAMF?
Limits:
(8 characters minus any extension)
Query Response:
Description:
The query will return the index number of the entry with name
has an index value of 1. If
(0) is returned.
Other Required Cmds:
None
Example:
N A M F ? V G A _ m 4 / / R e t u r n p o s i t i o n o f
/ / V G A _ m 4 i n d i r e c t o r y