Lim_notify (alarming), Lo_alm (alarming), Lo_lim (alarming) – National Instruments FP-3000 User Manual
Page 91: Lo_lo_alm (alarming), Lo_lo_lim (alarming), Lo_lo_pri (alarming), Lo_pri (alarming), Low_cut (i/o, option, scaling, tuning), Manufac_id (diagnostic)

Appendix B
Fieldbus Parameters
© National Instruments Corporation
B-13
FieldPoint FP-3000 User Manual
LIM_NOTIFY (Alarming)
A limit on the number of unconfirmed alarm/event notification messages
the device can have active at once. This must be less than or equal to
MAX_NOTIFY
.
LO_ALM (Alarming)
The current state of the low alarm, along with a time and date stamp.
LO_LIM (Alarming)
The limit, in PV units, beyond which the low limit alarm is considered
active.
LO_LO_ALM (Alarming)
The current state of the low-low alarm, along with a time and date stamp.
LO_LO_LIM (Alarming)
The limit, in PV units, beyond which the low-low limit alarm is considered
active.
LO_LO_PRI (Alarming)
The priority of the low-low limit alarm.
LO_PRI (Alarming)
The priority of the low limit alarm.
LOW_CUT (I/O, Option, Scaling, Tuning)
With an
L_TYPE
of Indirect Square Root, this can be used to establish a
floor (in percent of scale) for values from the transducer. Values below this
floor are considered to be zero. This feature must first be enabled by setting
Low Cutoff
in the
IO_OPTS
parameter.
MANUFAC_ID (Diagnostic)
The ID of the manufacturer of the device. For National Instruments devices,
it is 0x4E4943. The parameters
MANUFAC_ID
,
DEV_TYPE
,
DEV_REV
, and
DD_REV
are used in combination for a host tool to locate the Device
Description for this device.