LS8000-3 25-Pin Connector Pinout
Pin
Description
1
RS-232 Transmit (from LS8000-3 to host)
2
RS-232 Receive (from host to LS8000-3)
3
Phase A True – High Speed Output (RS-422 Drivers)
4
Phase A False – User Scaleable (5-24V Output)
5
Phase A False – High Speed Output (RS-422 Drivers)
6
Phase B True – User Scaleable (5-24V Output)
7
Phase B True – High Speed Output (RS-422 Drivers)
8
Phase B False – User Scaleable (5-24V Output)
9
Phase B False – High Speed Output (RS-422 Drivers)
10
Material Present Input (5-24V Input)
11
Signal Ground for Inputs/Outputs/Serial
12
Power Ground for 24V Input
13
Power Ground for 24V Input
14
Measurement Direction Input (5-24V Input)
15
Phase A True – User Scaleable (5-24V Output)
16
Laser Interlock (Connect to Signal Ground to Turn On Laser)
17
Shutter Control (Connect to Signal Ground to Open Shutter)
18
Length Reset Input (5-24V Input)
19
Signal Ground for Inputs/Outputs/Serial
20 User
V
IN
– Voltage input for Isolated Pulse Outputs (5 to 28V DC). The
voltage supplied will be the voltage level of the pulse outputs supplied by
the LS8000-3. If a Voltage is not supplied, the pulse outputs will be
approximately 4.5V.
21
Signal Ground for Inputs/Outputs/Serial
22
Index Pulse True - User Scaleable (5-24V Output)
23
Index Pulse False - User Scaleable (5-24V Output)
24 24V
Fused
Input
25 24V
Fused
Input