COMM LED Control
Indicates traffic on either RS232/RS422 or Ethernet. The light will turn blue
when the gauge is transmitting, turn orange when the gauge is receiving and
white when both transmit and receiving are happening at the same time.
The COMM light in conjunction with the Valid Light indicate the BAUD Rate
and control parameters for RS232/RS422 communication.
The COMM light will blink n times when the gauge is power up to indicate the
BAUD rate the gauge is set to.
The Valid light will be off during the power up sequence if the number of date
bits, parity and number of stop bit is set to 7-n-2 and on during power up
sequence when set to 8-n-1.