Qquality factor current value, Rreset internal length count – Laser beta lasermike LS8000-3 User Manual
Page 136

LaserSpeed 8000-3 Instruction Handbook
RS-232/RS-422 Communication
Part No. 93463 / Drawing No. 0921-01561
Page 136 of 221
Revision A (Sep 2007)
Q
Quality Factor Current Value
Q
Read Only Command
Response
Format:
is an integer from 0 to 15 indicating the current Quality Factor. A QF of 0 indicates that
the gauge is not currently making measurements, and a QF of 15 indicates that the gauge is
making excellent measurements.
This command returns the same value as you would normally receive from the TE data, or in
a response to the V command.
R
Reset Internal Length Count
R
or
Command
R0
This command causes the LS8000-3 to reset its internal length count, and to reset the Index
Pulse location to 0.