m/min x 1min/60sec x 10,000 pulses/m = 166,667 1/s ). The range is 0.001 to
100,000.
Measurement Units
Sets the units that the length and speed will be measured in. Available choices
are:
• Feet/sec
• Feet/minute
• Meters/sec
• Meters/min
• Inches/min
• Millimeters/min
• Millimeters/sec
• Yards/min
• Yards/sec
Length will be in m, feet, mm, in or yards, depending on the selected value.
User Update Rate
The core period determines how often the LaserSpeed sensors provide the
LaserTrak software with an updated speed and length measurement. Available
choices are Fast (4ms), Medium (32ms) and Slow (100ms) in the Standard
settings. For the advanced settings the available choices are 1 to 2000 ms in
1 ms increments.
The Recommended User Update Rate is 32 ms.
Output Configuration Parameters
Analog Zero Scale
Sets the speed equal to the analog zero scale. See the Analog Output Voltage
Section for details.
Analog Full Scale
Sets the speed equal to the analog full scale. See the Analog Output Voltage
section for details.
AFS Represents QF
When checked the gauge will output QF on the analog output as a 0 to 1 volt
signal. 0 - QF 0 and 1 volt –QF=15. When not checked the gauge will output
speed on the analog output as a 0 to 2 volt signal
Auxiliary Digital Output
Func.
Sets the operation of the Auxiliary output. The choices are show on the pop up
menu.
High Speed Pulse
Config
The high speed pulse output defaults to outputting pulse according to the High
Speed Pulse Rate set in the standard Configuration Tab. The pulse output can
be configured to output gauge status if the high speed pulse output is not
needed. The choices are shown in drop down menu.