Multichannel Systems Cardio2D Manual User Manual
Page 83
Cardio 2D+ Analyzer Software
79
Click "2D" tab in "Analyze" dialog.
The first up down box in this tab of the dialog defines the minimal number of cluster channels.
The "Min Cluster Channels" gives the minimum number of channels at which heartbeats have
to be detected to form a cluster. For example, for a MEA electrode layout with 60 electrodes
it could be 10, but if less than 60 electrodes are involved, it could be 5 or 2.
The second up down box defines the maximum propagation time of the heartbeat within
a cluster, for example the maximum duration between the first and the following detection
of a heartbeat on different electrodes.