Multichannel Systems Cardio2D Manual User Manual
Page 41
Cardio 2D Software
37
4.8
Stimulation with MEA2100-System
Click "Stimulate" icon
.
When working with Cardio2D and any type of MEA2100-Systems with integrated stimulus
generator STG the "Stimulate" dialog provides the possibility of stimulating via three different
stimulation patterns on user defined electrodes.
The dialog mirrors the layout of the currently used MEAs, in this case a MEA2100-2x60-System
with two slots for two 60 electrode MEAs. Use the tool bar thereunder to define the stimulation
sites.
Tool Bar
The tool bar in the stimulate dialog is divided into three sections: A position indicator,
controls for single stimulus patterns, and controls that affect all three stimulus patterns.
First section: The position indicator helps to navigate through the MEA electrodes.
It shows the position of the mouse on the left MEA 1 or on the right MEA 2
.
The numbering of MEA electrodes in the 8x8 grid follows the standard numbering scheme
for square grids: The first digit is the column number, and the second digit is the row number.
For example, electrode 23 is positioned in the second column and the third row.