Multichannel Systems Cardio2D Manual User Manual
Page 24

Cardio2D Manual
20
Tools
Menu to activate the offset correction, if necessary. Enable the option "Offset Correction" and the
following dialog appears.
An offset correction may be necessary, if you observe disturbing voltage offset on the input
channels. Click the check box "Correct Offset" and Cardio2D takes 100 ms of the recorded data in
the moment when the check box is selected to calculate the DC offset. The mean of this 100 ms
sweep is subtracted from the recorded data as long as the "Correct Offset" check box is activated.
The offset correction works separately for each available channel. Click the "Update" button to
update the calculation of the offset correction. This option is available when Cardio2D is running.
Settings
Menu to define the data acquisition device and to determine the path for the data and log files.