Sending the configuration file, Viewing version information, Warm booting the module – ProSoft Technology MVI69-101S User Manual
Page 62: Viewing data type setup

MVI69-101S ♦ CompactLogix or MicroLogix Platform
Diagnostics and Troubleshooting
IEC 60870-5-101 Slave Communication Module
Page 62 of 149
ProSoft Technology, Inc.
March 16, 2009
Sending the Configuration File
Press [S] to upload (send) an updated configuration file to the module. For more
information on receiving and sending configuration files, please see Uploading
and Downloading the Configuration File.
Viewing Version Information
Press [V] to view Version information for the module.
Use this command to view the current version of the software for the module, as
well as other important values. You may be asked to provide this information
when calling for technical support on the product.
Values at the bottom of the display are important in determining module
operation. The Program Scan Counter value is incremented each time a
module's program cycle is complete.
Tip: Repeat this command at one-second intervals to determine the frequency of program
execution.
Warm Booting the Module
Caution: Some of the commands available to you from this menu are designed for advanced
debugging and system testing only, and can cause the module to stop communicating with the
processor or with other devices, resulting in potential data loss or other failures. Only use these
commands if you are specifically directed to do so by ProSoft Technology Technical Support staff.
Some of these command keys are not listed on the menu, but are active nevertheless. Please be
careful when pressing keys so that you do not accidentally execute an unwanted command.
Press [W] from the Main Menu to warm boot (restart) the module. This command
will cause the program to exit and reload, refreshing configuration parameters
that must be set on program initialization. Only use this command if you must
force the module to re-boot.
Viewing Data Type Setup
To view setup information for each data type, press the matching key from the
Menu.
Key
Data Type
Screen Example
1
M_SP_NA
2
M_DP_NA