KLP-DEV 041213
3-15
FIGURE 3-3. SETTING THE OUTPUT AND TAKING A MEASUREMENT EXAMPLE, WRITTEN IN VISUAL
BASIC (SHEET 1 OF 2)
'/*~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~
' This program demonstrates how to set the output voltage and current
' measure the voltage
' using the IVI-COM driver.
'/*~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~
Option Explicit
Public driver As KepcoDCPwr
Public outputPtr As IKLPOutput
Public protectionPtr As IKLPProtection
Public measurementPtr As IKLPMeasurement
Private Sub cmdExit_Click()
Unload Me
End Sub
Private Sub cmdStart_Click()
 
 Set driver = New KepcoDCPwr
 
 On Error GoTo ErrorHandler
 
 
 ' initialize the driver - note ID device is true
 driver.Initialize "GPIB0::6::INSTR", _
 True, _
 True, _
 "Cache=true,InterchangeCheck=false,QueryInstrStatus=true,Simu-
late=false"
 
 Dim result As Boolean
 result = driver.Initialized
 
 
 Set outputPtr = driver.Outputs.Item(driver.Outputs)
 Set protectionPtr = driver.Protections.Item(driver.Protections)
 Set measurementPtr = driver.Measurements.Item(driver.Measurements)
 
 ' set voltage level
 outputPtr.VoltageLevel 75
 
 ' enable OV protection and set OV level
 protectionPtr.ConfigureOVP 15
 
 
 ' set current level
 outputPtr.CurrentLimit 10
 
 ' enable the output
 outputPtr.Enabled = True
 
 Delay 10
 
 ' Measure the voltage
 Dim measVoltage As Double
 measVoltage = measurementPtr.Measure(KepcoDCPwrMeasurementVoltage)
 