Table 3-2. factory defaults, Factory defaults -4 – KEPCO KLN 750W Series, Firmware Version 1.60 to 1.6x User Manual
Page 40

3-4
KLN 750W 032614
4. When the self-test is complete, the Voltage display shows the last voltage setpoint value and
the Current display shows the last current setpoint value. All the front panel keys are func-
tional and the remote-controlled functions can be enabled.
• If an error occurs, the red
(alarm) LED goes on. Refer to PAR. 3.3.45 to view the
error code.
5. After a successful self test, the factory default conditions upon power up are as shown in
Table 3-2.
TABLE 3-2. FACTORY DEFAULTS
Parameter
Factory Default
Output V
10% of rated voltage (0V for E- or G- suffix models)
Output C
10% of rated current (0A for E- or G- suffix models)
OVP
110% of rated Voltage
OCP
110% of rated Current
Output: On/off
OFF
Power On: LAST/OFF
OFF
RAMP UP time
0.1
RAMP down time
0
Undervoltage Limit
0
BEEP: On/Off
On
I/O Select
RS 485
RS 485 Baud rate
115200
GPIB address
07
RS 485 address
007
Parallel/Series: PM/PS/SM/SS
PM
External Control of output: ON/OFF
OFF
Voltage Control: LOC/EXT
LOC
Current Control: LOC/EXT
LOC
Front panel lockout
OFF
Voltage Set: Direct/Enter
Direct
DHCP: On/Off
On
SYST:COMM:LAN:GATE
DHCP On: from DHCP server
DHCP Off: 0.0.0.0
SYST:COMM:LAN:IP
DHCP On: From DHCP server
DHCP Off: 192.168.0.100
SYST:COMM:LAN:IPAD
192.168.0.100
SYST:COMM:LAN:SMAS
255.255.255.0
SYST:COMM:LAN:MAC
70-46-42-XX-XX-XX
SYST:COMM:LAN:TEL:PORT 8003
NOTES:
(1) Parameters changed via front panel are saved when unit is powered off, and will be restored upon power-up.
Parameters changed via SCPI command are not saved when unit is turned off and are restored to the last
value entered from front panel when power is cycled on.
(2) *RST resets the following: Output to OFF, Output V to 10% of rated voltage, Output A to 10% of rated current,
OVP to 110% of rated voltage, OCP to 110% of rated current, Undervoltage Limit to 0, and front panel lockout
to OFF. All other parameters shown above are restored to the last value entered from front panel.
(3) Cycling power resets the following: Output to setting determined by Power On setting (see PAR. 3.3.9), Under-
voltage Limit to 0 and front panel lockout to OFF. All other parameters shown above are restored to the last
value entered from front panel.