Grand total reading ( grtotal ), Toggle lower display parameter – Precision Digital PD6830 User Manual
Page 91

PD6830 Pulse Input Rate/Totalizer Instruction
Manual
91
Grand Total Reading (GrTOTAL)
The grand total is a separate total that is not reset when the total is reset. This allows
the complete total to be tracked by the grand total, while individual batch, or daily
totals are reset regularly.
To display the grand total, press the Up/Display button. The display will read
GRTOTAL
, and the GT symbol will appear indicating the grand total is being displayed
on the bottom display. After 10 seconds, the bottom display will return to showing
total. To lock the grand total on the display, press Enter. Pressing Menu at any time
will return to normal run mode.
Note: If the Display menu has been setup to display the grand total on the bottom
display, pressing the Up/Display
button will display the maximum and minimum
readings followed by the total.
Toggle Lower Display Parameter
The maximum and minimum (peak & valley) readings reached by the rate are stored
in the meter since the last reset or power-up. The meter shows MAXIMUM or MINIMUM
to differentiate between run mode and max/min display. The previous total and grand
total value reached prior to being reset are also stored.
To display the maximum and minimum readings or the previous total and grand total
use Up/Display button to cycle the bottom display. Maximum and minimum are
displayed after the grand total.
Press the Enter button to remain in Max/Min/Previous display mode. If Enter is not
pressed, the Max/Min/Previous display readings will time out after ten seconds. The
meter will return to display the actual reading. Pressing Menu at any time will return
to normal run mode.