Commissioning and preparing to measure

Device setup

Operating Instructions

91

Select signal period

The signal period is based on the type of 1 V

PP

or TTL linear encoder connected to the

channel. The signal period is used to convert the count returned from the encoder to
a linear position. Refer to the documentation provided with the device for the correct
signal period setting.

To select a signal period:

Click the signal period drop-down arrow and select the desired value.

Click the «OK» button.

A message verifying the action is displayed in the message area.

1 V

pp

encoder

Signal
period

Reference
marks

ST 128x

20 µm

Single

ST 308x

20 µm

Single

LS 388C

20 µm

Coded/1000

LS 688C

20 µm

Coded/1000

LS 187
LS 187C

20 µm
20 µm

Single
Coded/1000

LS 487
LS 487C

20 µm
20 µm

Single
Coded/1000

LB 382C

40 µm

Coded/2000

LF 183
LF 183C

4 µm
4 µm

Single
Coded/5000

LF 483
LF 483C

4 µm
4 µm

Single
Coded/5000

TTL Encoder

Resolution

Signal
period

Interpolation
factor

Reference
marks

LS 177
LS 477

1 µm
0.5 µm
0.25 µm

20 µm
20 µm
20 µm

5-fold
10-fold
20-fold

Single
Single
Single

LS 177C
LS 477C

1 µm
0.5 µm
0.25 µm

20 µm
20 µm
20 µm

5-fold
10-fold
20-fold

Coded/1000
Coded/1000
Coded/1000

LS 328C
LS 628C

5 µm

20 µm

n/a

Coded/1000

LS 378C

1 µm
0.5 µm
0.25 µm

20 µm
20 µm
20 µm

5-fold
10-fold
20-fold

Coded/1000
Coded/1000
Coded/1000