beautypg.com

HEIDENHAIN MSE 1000 User Manual

Page 82

background image

Commissioning and preparing to measure

Module setup

82

Operating Instructions

To view the LVDT Teach resolutions screen:

Click the «Setup» button.

Select an LVDT module from the Tree or the Module view.

Click the «Calibration» button.

Click the «OK» button to the right of Teach resolutions.

The LVDT Teach resolutions screen for the selected module appears.

Teach resolution for a
single sensor

To teach the resolution for a single sensor:

Use a gage block to position the sensor so that the Sensor output (V) value shows
as close to 0.0 as possible.

Move the sensor to the minimum mechanical position

Click inside the Min (mm) or Min (in) field and enter the distance that was traveled
from the 0.0 position.

Click the «OK» button to the right of the Min (mm) or Min (in) field.

Use a gage block to position the sensor so that the Sensor output (V) value shows
as close to 0.0 as possible.

Move the sensor to the maximum mechanical position.

Enter the distance that was traveled from the 0.0 position in the Max (mm) or Max
(in) field.

Click the «OK» button to the right of the Max (mm) or Max (in) field.

The resolution that will be used to convert from volts to millimeters or inches will
be displayed in the Res. (mm/V) or Res. (in/V) field.

The Calibration timestamp is updated.

Teach resolution for all
sensors

To teach the resolution for all sensors:

Use gage blocks to position all of the sensors so that the Sensor output (V) values
show as close to 0.0 as possible.

Move all of the sensors to their minimum mechanical positions.

Click inside each Min (mm) or Min (in) field and enter the distance that was
traveled from the 0.0 position.

Click the «SetAllMin» button.

Move all of the sensors to their maximum mechanical positions.

Click inside each Max (mm) or Max (in) field and enter the distance that was
traveled from the 0.0 position.

Click the «SetAllMax» button.

The resolutions that will be used to convert from volts to millimeters or inches will
be displayed in the Res. (mm/V) or Res. (in/V) fields.

The Calibration timestamps are updated.