Troubleshooting

Module warnings and errors

Operating Instructions

145

Warning or error

Cause

Corrective action

Temperature is below the

minimum error tolerance
Power LED:
Flashing red and
green at a rate of once every three
seconds.

The module's CPU temperature is
< -5 °C.

Verify that Operating Temperature

specifications are being met.

Refer to "Module specifications",

page 158.

Temperature exceeds the

maximum error tolerance
Power LED:
Flashing red and
green at a rate of once every three
seconds.

The module's CPU temperature
is > 110 °C.

Verify that Operating Temperature

specifications are being met.

Refer to "Module specifications",

page 158.

Turn the power off.

Allow the module to cool off.

Verify that adequate ventilation is

available for the module to remain

within operating temperatures.

If the error persists contact

HEIDENHAIN technical support for

assistance.

The module non-volatile memory

is using default data
Power LED:
Flashes red and green
at a rate of once every ten seconds.

A module had to use default
configuration parameters instead
of the normal configuration
parameters.

Clearing the error will remove it
from the Error screen. The error will
appear again on the next power-up
of the module.

Clear the error in the Error screen.

Contact HEIDENHAIN technical

support for assistance.

The module non-volatile memory

has invalid data, using backup

A module had to use its backup
configuration parameters instead
of the normal configuration
parameters.

Clearing the error will remove it
from the Error screen. The error will
appear again on the next power-
up of the module if the backup
parameters were not successfully
copied.

Verify that the power supply is

running at an acceptable level.

Verify that the power supply

does not have too many modules

connected causing an over-current.

If the error persists, contact

HEIDENHAIN technical support for

assistance.