Troubleshooting
Application errors
142
Operating Instructions
Error
Cause
Corrective action
The modules failed to be set
Use DHCP Addressing failed.
Power cycle the module chain.
Refer to "Power cycling", page 33.
Perform a Reconnect or Broadcast.
Set the module chain to Use DHCP
Addressing.
The image chosen for programming
is invalid
An invalid file was selected to
program the bootloader or firmware.
Select a valid MSEbootloader.dat or
MSEfirmware.dat file.
The checksum failed
The MSEbootloader.dat or
MSEfirmware.dat file has become
corrupted.
Select a valid MSEbootloader.dat or
MSEfirmware.dat file.
The bootloader and firmware
cannot be updated while using
DHCP
The MSEbootloader.dat or
MSEfirmware.dat failed to program
since the module is in DHCP mode.
Set the chain to static IP
addressing and retry.
All modules are already the desired
version
Programming the firmware for
All modules failed because the
firmware version is the same.
Modules can be programmed
individually without a check for the
version.
Program modules individually.
Password incorrect
An invalid password has been
entered for Supervisor mode.
Enter the correct password.
The IP addresses could not be set
because of duplicates
The Use Static Addressing button
in the networking screen failed
to set the modules to their static
IP addresses because there are
duplicates.
Perform a broadcast and then
allow the DHCP addresses to be
set as static IP addresses when
prompted.
or
Set the static IP addresses
individually using the Set Static
button in the networking screen.