Troubleshooting

Application errors

140

Operating Instructions

Error

Cause

Corrective action

MSEsetup was not shutdown

properly

The workstation was powered off or
the application was closed through
the operating system instead of the
«Close» button.

MSEsetup will only need to
recover from this error if the
System Configuration or Module
Configuration file has been
corrupted by the improper
shutdown.

Restore defaults

or

Load previously saved System

Configuration and Module

Configuration files. Refer to "File

options", page 57.

If MSEsetup will not open:

Delete the System Configuration

and Module Configuration files.

Refer to File locations, page 34 for

configuration file locations.

Open MSEsetup.

Setting the IP address failed

The Set Static operation failed.

Power cycle the module chain.

Refer to "Power cycling", page 33.

Perform a Broadcast.

The module chain needs to be

reconnected

The MSE 1000 modules were
powered off while MSEsetup is
connected.

Perform a Reconnect.

Error programming module

Programming the firmware
or bootloader for a module
failed. The Module label and the
Communication error will also be
displayed.

Power cycle the module chain.

Refer to "Power cycling", page 33.

Perform a Broadcast.

It may be necessary to power

cycle twice in a row to clear

the module’s state machine if

programming fails.

Warning, error returned from

module

An error occurred while
communicating to a module.
MSEsetup will retry 5 times before
the

Lost communication to the MSE

1000 error will occur.

Error recovery is only needed if the
Lost communication to the MSE
1000
error occurs. Refer to this
error above.