HEIDENHAIN MSE 1000 User Manual
Page 140

Troubleshooting
Application errors
140
Operating Instructions
Error
Cause
Corrective action
MSEsetup was not shutdown
properly
The workstation was powered off or
the application was closed through
the operating system instead of the
«Close» button.
MSEsetup will only need to
recover from this error if the
System Configuration or Module
Configuration file has been
corrupted by the improper
shutdown.
Restore defaults
or
Load previously saved System
Configuration and Module
Configuration files. Refer to "File
If MSEsetup will not open:
Delete the System Configuration
and Module Configuration files.
Refer to File locations, page 34 for
configuration file locations.
Open MSEsetup.
Setting the IP address failed
The Set Static operation failed.
Power cycle the module chain.
Refer to "Power cycling", page 33.
Perform a Broadcast.
The module chain needs to be
reconnected
The MSE 1000 modules were
powered off while MSEsetup is
connected.
Perform a Reconnect.
Error programming module
Programming the firmware
or bootloader for a module
failed. The Module label and the
Communication error will also be
displayed.
Power cycle the module chain.
Refer to "Power cycling", page 33.
Perform a Broadcast.
It may be necessary to power
cycle twice in a row to clear
the module’s state machine if
programming fails.
Warning, error returned from
module
An error occurred while
communicating to a module.
MSEsetup will retry 5 times before
the
Lost communication to the MSE
1000 error will occur.
Error recovery is only needed if the
Lost communication to the MSE
1000 error occurs. Refer to this
error above.