Troubleshooting

Communication errors

146

Operating Instructions

12.5

Communication errors

The error

Lost communication to the MSE 1000 is sent to the Error screen for communication errors between

MSEsetup and MSE 1000 modules. The actual error as well as the module that had the error and the channel, if
applicable, will be shown in the Log screen.

If the troubleshooting procedures described in this section are unsuccessful, refer to "Network troubleshooting",
page 151
for additional information.

Error

Cause

Corrective action

Response timeout

A command was sent to a module
and failed to obtain a response
within the timeout period. The
default timeout period is 200 ms.

Check the power and network

connection to the modules.

Power cycle the module chain.

Refer to "Power cycling", page 33.

Perform a reconnect.

Broadcast failed to get a response

A broadcast returned with no
modules found.

Close MSEsetup.

Restart MSEsetup.

Verify that the workstation IP is on

the correct domain.

Verify that communication to a

module is working with a ping

request from the operating system.

Perform a broadcast when

communication is verified.

The binding of the UDP socket

failed

A broadcast could not be started
because the socket failed to bind.
This is usually because another
application is using the socket and
port.

Close MSEsetup.

Restart MSEsetup.

Verify that the workstation IP is on

the correct domain.

Verify that communication to a

module is working with a ping

request from the operating system.

Change the port from the

networking screen if another

application is using the port.

Perform a broadcast when

communication is verified.