9

Switch-On, Traversing the Reference Marks

ENT...CL

Turn on power (switch located on rear panel).

“REF” in status display starts blinking.

Confirm reference traverse mode. “REF”

remains on continuously. Decimal points start

blinking.

Cross over the reference marks in all axes in any

sequence. Each axis display becomes active

when its reference mark is crossed over.

0 è 1

ENT

Crossing over the reference marks stores the most recently defined

assignment of display values to axis slide positions for datum points 1

and 2 in nonvolatile memory.

Note that if you choose not to traverse the reference marks (by

clearing the dialog ENT ... CL with the CL key), this relationship will be

lost if the power is interrupted or when the unit is switched off.

Switch-On,

Traversing

the

Reference

Marks

If you wish to use multipoint axis error compensation you

must traverse the reference marks (see “Multipoint axis error

compensation”)!