Measured Value Output

Measured values can be output over the RS-232-C/V.24 interface. This

can be done with the following functions:
Probing with the KT Edge Finder

Contact input on X10

Pulse input on X10

CTRL B via the RS-232-C interface

Code letters with the measured value
Parameter P97 allows you to select a code letter to be output together

with the measured value when using Probe, Contact or Pulse. The

decimal number you enter in the parameter is the ASCII character

number in the ASCII table. If you enter 0, no code letter will be output.

The code letter enables you to recognize whether the measured value

was generated with CTRL B or with an external signal.

Axis designation for measured value output
Parameter P49 allows you to enter an axis designation for each

measured value that is output. The decimal number you enter in the

parameter is the ASCII character number in the ASCII table. If you enter

0, no axis designation will be output.

Example of measured value output:
Parameter settings:

P49.1 =

88

(“X”)

P49.2 =

89

(“Y”)

P49.3 =

90

(“Z”)

P51

=

0

(no line feeds)

P97

=

69

(“E”)

Output:

E (CR)(LF)

X=...(CR)(LF)

Y=...(CR)(LF)

Z=...(CR)(LF)

Measured

V

alue

O

utput