beautypg.com

Error messages – HEIDENHAIN ND 720 v.3 User Manual

Page 30

background image

30

To erase error messages:

After you have removed the cause of error:

➤ Press the CL key.

Error

Messages

Error Messages

Message

Cause and Effect

RS TOO FAST

Time interval between two commands

too short.

SIGNAL X

Encoder signal is too weak,

e.g. when an encoder is contaminated.

PROB. ERROR

Before touching off on the workpiece,

the tool must move by a distance of at

least 0.2 mm.

DSR MISSING

The connected device is not

transmitting a DSR signal.

ERR. REF. X

The spacing of the reference marks as

defined in P43 is not the same as the

actual spacing.

FORMAT. ERR.

Data format, baud rate, etc. do not

match.

FRQ. X

The input frequency for this encoder

input is too high. This can occur when

the scale is moved too fast.

Message

Cause and Effect

ERR. MEMORY

Checksum error: Check the

datum, operating parameters and

compensation values for

multipoint axis error

compensation. If the error recurs,

contact your service agency!

REC. ERROR

Error during reception of

parameter lists or

compensation-value lists.

Bateil1.pm6

07.11.2001, 10:36

30