Operating parameter list – HEIDENHAIN ND 720 v.3 User Manual
Page 37

37
Operating
Parameters
List of operating parameters
P00 CODE Enter code number
9 51 48:
Change protected operating parameters
66 55 44:
Display the software version (in the X axis)
Display the date of release (in the Y axis)
10 52 96:
Multipoint axis error compensation
48 61 53:
Input and output of parameter lists and
compensation-value lists
P01 Unit of measure
1)
Display in millimeters
MM
Display in inches
INCH
P03.1 to P03.3 Radius/diameter display
1)
Display position value as radius
RADIUS
Display position value as diameter
DIAMETER
P11 Activate scaling factor
1)
Active
SCALING ON
Not active
SCALING OFF
P12.1 to P12.3 Define scaling factor
1)
Enter a scaling factor separately for each axis:
Entry value > 1: workpiece will grow
Entry value = 1: workpiece will remain the same size
Entry value < 1: workpiece will shrink
Input range:
0.100000 to 9.999999
Default setting:
1.000000
P25 Stylus diameter
1)
Input range (mm):
0.000 to 999.999
Default setting:
6
P26 Stylus length
1)
Input range (mm):
0.000 to 999.999
P30.1 to P30.3 Counting direction
Positive counting direction with
positive direction of traverse
DIRECT. POS
Negative counting direction with
positive direction of traverse
DIRECT. NEG
P31.1 to P31.3 Signal period of the encoder
Input range:
0.00000001 to 99999.9999 µm
Default setting:
20 µm
P33.1 to P33.3 Counting mode
0 - 1 - 2 - 3 - 4 - 5 - 6 - 7 - 8 - 9
0 - 2 - 4 - 6 - 8
0 - 5
P38.1 to P38.3 Decimal places
2 / 3 / 4 / 5 / 6 / 7 / 8
1)
User parameter
Bateil2.pm6
27.03.2002, 12:09
37