Delay times with data output
Duration of the latch signal:
t
e
³ 4 µs
Storage delay:
t
1
£ 4.5 ms
Data output after:
t
2
£ 50 ms
Regeneration time:
t
3
³ 0
Duration of data output in seconds:
209 number of axes + 11 number of line feeds
t
D
=
Baud rate
Measured value output when probing
Parameter P96 allows you to activate measured value output when
probing with the KT edge finder. The edge finder is connected to
D-sub input X10.
Whenever the Probe Edge function is used, your display unit outputs
the position of the edge in the selected axis and the actual positions of
the other axes over the TXD line of the RS-232-C/V.24 interface.
Whenever the Probe Midpoint function is used, your display unit
outputs the calculated midpoint in the selected axis and the actual
positions of the other axes.
Measured value output with CTRL B is disabled when a probing
function is active.
Next possible signal for measured value output: t
E
= t
1
+ t
2
+ t
D
+ t
3
[s]
t2
t1
t3
tD
te
te
Measured
V
alue
O
utput