Output
Structure
of
Parameter
Lists
ND 720 parameter list: (factory default setting)
Parameter list
Description
*
Starting character (*);
ND-720
Type of display unit
P01
MM =
0
MM = 0; INCH = 1;
P03.1
RADIUS =
0
X axis: RADIUS = 0; DIAMETER = 1;
P03.2
RADIUS =
0
Y axis: RADIUS = 0; DIAMETER = 1;
P11
SCALING OFF =
0
SCALING OFF = 0; ON = 1;
P12.1
SCL =
1.000000
X axis: SCALING FACTOR = 1.000000; (value input)
P12.2
SCL =
1.000000
Y axis: SCALING FACTOR = 1.000000; (value input)
P25
PROBE DIA. =
+ 6.000
PROBE DIAMETER = 6.000 mm; (value input)
P30.1
DIRECT.POS =
0
X axis: DIRECTION POS. = 0; NEG. = 1;
P30.2
DIRECT.POS =
0
Y axis: DIRECTION POS. = 0; NEG. = 1;
P31.1
S.PER. =
20
X axis: SIGNAL PERIOD = 20µm; (value input)
P31.2
S.PER. =
20
Y axis: SIGNAL PERIOD = 20µm; (value input);
P33.1
COUNT 0-1 =
1
X axis: COUNT 0-1 = 1; 0-2 = 2; 0-5 = 5;
P33.2
COUNT 0-1 =
1
Y axis: COUNT 0-1 = 1; 0-2 = 2; 0-5 = 5;
P38.1
DP POS. 3 =
3
X axis: DECIMAL PLACES 3 (range: 1-8)
P38.2
DP POS. 3 =
3
Y axis: DECIMAL PLACES 3 (range: 1-8)
P40.1
COMP. OFF =
0
X axis: COMPENSATION OFF = 0; LIN = 1; MULTI = 2;
P40.2
COMP. OFF =
0
Y axis: COMPENSATION OFF = 0; LIN = 1; MULTI = 2;
P41.1
LINEAR CMP. =
+ 0
X axis: LINEAR COMPENSATION = 0 µm/m (value input)
P41.2
LINEAR CMP. =
+ 0
Y axis: LINEAR COMPENSATION = 0 µm/m (value input)
P42.1
BKLASH =
+ 0.000
X axis: BACKLASH COMPENSATION 0.000 mm (value input)
P42.2
BKLASH =
+ 0.000
Y axis: BACKLASH COMPENSATION 0.000 mm (value input)
P43.1
1000 SP =
1000
X axis: ONE REF.M. = 0; 500; 1000; 2000; 5000 SP;
P43.2
1000 SP =
1000
Y axis: ONE REF.M. = 0; 500; 1000; 2000; 5000 SP;
P44.1
REF. ON =
1
X axis: REF.ON = 1; REF. OFF = 0;
P44.2
REF. ON =
1
Y axis: REF.ON = 1; REF. OFF = 0;
P45.1
ALARM ON =
1
X axis: ALARM ON = 1; ALARM OFF = 0;
P45.2
ALARM ON =
1
X axis: ALARM ON = 1; ALARM OFF = 0;