The table below gathers the characteristics of all clock inputs/outputs:

External Reference Input

Input Voltage Level

1 – 3.3 Volts peak-to-peak (AC-coupled)

Frequency Range

0 – 100 MHz.

External Sampling Clock Input

Input Voltage Level

1.5 – 3.3 Volts peak-to-peak (AC-coupled)

Input Format

Single-ended.

Frequency range

10-500 MHz

External Sampling Clock Output

Output Voltage Level

0-2.4 Volts fixed amplitude

Output Format

LVTTL

External Trigger Inputs

Input Voltage Level

1.2-2.5 Volts peak-to-peak.

Format

DC-coupled and Single-ended. Protected by 2

clamping diodes.

Format

LVCMOS

Figure 11 - Clock Architecture Main Characteristics.


Below are shown some captures of the external clock output:

Figure 12 - External clock output (98.304MHz

– Divided by 5 output)