Fig.5-80
:Cursor time measurement
Cursor measure position menu as below:
Function
Display
Instruction
M1-M2
Position
The position difference between two cursors in memory area.
Cursor 1
Position
The position of cursor 1 corresponds to trigger in memory area.
Cursor 2
Position
The position of cursor 2 corresponds to trigger in memory area.
Follow up below steps to measure data width of pulse with cursor measurement
1. Press "9(CURSOR)" and Cursor menu appears.
2. Press "F1" till cursor display as "Position". Two purples vertical lines display in panel
and refer to Cursor 1 and Cursor 2 separately.
3. Turn "CH1 POSITION" knob and set Cursor 1 position in falling edge of CH00
4. Turn "CH2 POSITION" knob and set Cursor 2 position in next falling edge of CH00.
5. Now the position difference between two cursors display in "F2" is "-10" which means
there is 10 acquired data between two cursors. The position of Cursor 1 corresponds to
trigger in memory position display in "F3" as "2097158". The position of Cursor 2
corresponds to trigger in memory position display in "F4" as "2057168".