beautypg.com

How to set trigger system – OWON MSO Series User Manual

Page 76

background image

71

3. Press "F2" till storage depth display as "Deep Memory".
4. Press "F3" till digital filter display as "None".
Then sample system setting finished (refer to

Fig.5-53

).

Fig.5-53: Sampling setting

How to set trigger system

Trigger system including "Trigger level" adjust knob, "Trigger menu", "SET 50%",
"Set to Zero". Trig level adjust knob is to adjust trigger position. Since the digital storage
depth is fixed so the proportion of storage data between pre-trigger points and re-trigger
ones depend on trigger position. Trigger mode including: Edge trigger, bus trigger, pattern
trigger, sequential queue trigger, distributed queue trigger.
1. Edge trigger: make a channel as trigger source and set rising edge, falling edge or
either edge as trigger condition to generate trigger. Edge trigger menu refer to

Fig.5-54

.

Fig.5-54

: Edge trigger menu

Edge trigger function menu as below: