beautypg.com

OWON MSO Series User Manual

Page 88

background image

83

Fig.5-70

:Bus menu


BUS setting menu as below:

Function

Setting

Instruction

BUS

BUS0~BUS3

Choose BUS for operating

Channel

CH0F~CH00

1X111111

11111111

Choose any channel among CH00~CH0F
BUS channel complex indication: 1 for include; X
for exclude

CH0F~

CH00

Include
Exclude

The bus selected includes this channel
The bus selected excludes this channel.

Code

Type

HEX

DEC

All of the code type and measured value are HEX
All of the code type and measured value are DEC


For example, to set BUS0 as BUS to include CH00,CH01,CH02,CH03 and code as
hex system. Follow up as below:
1. Press "0(BUS)" to and BUS menu appears
2. Press "F1" till BUS display as "BUS0"
3. Press "F2" or turn "CH1 Volts/div" knob till channel display as "CH00"
4. Press "F3" and set CH00 as "Include". Then Channel CH00 refer to display of "1"
5. Repeat the operation of steps 2. 3 and set CH01, CH02, CH03 as "Include" and other
channel as "exclude". Then "CH SEL" display as "XXXXXXXXXXXX1111". BUS
setting finished.
6. Press "F4" and set code type as "HEX"
BUS setting is finished (refer to

Fig.5-71

).